|
No signal |
Assay set up incorrectly or used incorrect reagents |
Review protocol. Repeat assay using a positive control |
|
|
Incorrect secondary antibody used |
Retrace steps. Repeat assay using the correct secondary antibody |
|
|
Not enough antibody used |
Increase concentration of the primary and/or secondary antibody |
|
|
Detection reagent too old or contaminated |
Use fresh detection reagents |
|
|
Antigen not coated properly |
Try longer coating times, different coating buffers, or avidin plates with biotinylated antigen |
|
|
Plate reader has the wrong settings |
Check plate reader for wavelength, filters, gain etc. |
|
|
Antibody stored at 4°C for several weeks or subjected to repeated freeze/thaw cycles |
Use a fresh aliquot of antibody that has been stored at -20°C or below |
|
Weak signal |
Insufficient amount of antigen was coated to microtiter plate |
Use more antigen for coating or very coating buffer |
|
|
Not enough antibody used |
Increase concentration of the primary and/or secondary antibody. Optimize antibody concentrations for your assay |
|
|
Detection reagent too old, contaminated, or used at the wrong pH |
Use fresh detection reagents at the correct pH |
|
|
Detection reagent too dilute |
Use a higher concentration of detection reagent |
|
|
Plate reader has the wrong settings |
Check plate reader for wavelength, filters, gain etc |
|
|
Incubation temperature too low |
Optimize the incubation temperature for your assay. Reagents should be at room temperature before beginning the assay |
|
High background signal |
Too much antibody used |
Reduce the concentration of the primary and/or secondary antibody. Optimize antibody concentrations for your assay |
|
|
Non-specific antibody binding |
Use a suitable blocking buffer or use an affinity-purified antibody |
|
|
Too much detection reagent used |
Repeat assay with a higher dilution of detection reagent |
|
|
Too few washing cycles |
Increase the number of washing cycles |
|
|
Contaminated blocking agent |
Use fresh blocking agent |
|
|
Wrong concentration of blocking agent |
Check the concentration of blocking agent in the recommended protocol |
|
|
Presence of blocking buffer interferes with antibody binding |
Wash off blocking buffer before adding antibody |
|
|
Reaction not stopped |
Use stop solution to prevent overdevelopment |
|
|
Plate left too long before reading |
Start taking measurements shortly after the addition of detection reagent |
|
|
Wrong settings on the plate reader |
Check settings and adjust as needed |
|
|
Insufficient amount of Tween® in the buffers |
Use PBS containing 0.05% Tween® |
|
|
Incubation with substrate carried out in the light |
Perform substrate incubation in the dark |
|
|
Incubation temperature too high |
Optimize the incubation temperature for your assay |
|
|
Plates stacked during incubations leading to uneven temperature distribution |
Avoid stacking plates |
|
|
Pipetting errors |
Calibrate pipettes so that they dispense the correct volumes |
|
|
Reagents were not mixed properly |
Before pipetting solutions into wells, make sure all reagents and samples have been throughly mixed |
|
|
Inconsistent washing of wells |
Take precautions to reduce variability of washes |
|
|
Uneven evaporation of solution from wells during incubation |
Always incubate with a lid on the plate |
|
|
Interference caused by dirt on the bottom of the plate |
Clean the plate carefully and reread |
|
Slow color development |
Incubation temperature is wrong |
Ensure plates and reagents are kept at room temperature |
|
|
Contaminated solutions |
Make fresh solutions |
|
|
Detection reagent too old, contaminated or used at the wrong pH |
Use fresh detection reagents at the correct pH |