Biological Controls

Fig.1. Use of a positive control. Human peripheral blood was stained for IL-8 (MCA2709F) and CD11b A700 (MCA551A700). Cells were gated on mononuclear cells, the monocytes expressed IL-8 following stimulation with LPS for 5 hours.

Fig.1. Use of a positive control. Human peripheral blood was stained for IL-8 (MCA2709F) and CD11b A700 (MCA551A700). Cells were gated on mononuclear cells, the monocytes expressed IL-8 following stimulation with LPS for 5 hours.

Biological controls are crucial to determine whether your experiment has worked effectively. 

In addition to staining and isotype controls, you should also consider biological controls that will enable you to determine staining specificity and experimental limitations.

These controls are important for all staining but are more important for intracellular staining, which can have higher background fluorescence. Controls include known negative samples and known positive samples.

Examples of these are:

  • Cells known from the literature to either express or lack expression of the antigen of interest
  • Cells where the antigen has been knocked down or out, using RNAi or CRISPR technology to produce a negative cell
  • Cells which have been transfected and the antigen is being over expressed to ensure positive staining

For some experiments, such as cytokine release measurement, an unstimulated and fully stimulated sample is important to include to determine both positive results and the dynamic range of fluorescence staining.